System integrators evaluating a pH sensor spend most of their time on questions that have nothing to do with the glass bulb: which register holds the value, what data format it uses, how the station address is changed, and how calibration is triggered from a controller. Daruifuno has published the DPH890 user manual to answer those questions before a purchase order is raised.
The DPH890 is a Modbus RTU server device on an RS485 bus. The manual lists every parameter with its address, data format, permitted range and initial value, so commissioning does not depend on trial and error:
The three calibration point values are not fixed. Point 1 can be set between 3.5 and 4.51 pH, point 2 between 6.36 and 7.5 pH and point 3 between 8.68 and 10.5 pH, so a plant can calibrate with the standard solutions it already holds rather than ordering a specific set. Zero point, slope 1, slope 2, main measurement offset and temperature offset are all writable within published ranges, which gives maintenance teams a documented way to compensate for ageing before replacing an electrode.
| Parameter | Specification |
|---|---|
| Measuring principle | Glass electrode method |
| Measuring range | 0.00~14.00 pH; 0.0~50.0 °C |
| Resolution / accuracy | 0.01 pH / 0.1 °C; 0.1% FS or ±0.02 pH |
| Temperature compensation | Automatic or manual |
| Communication | RS485 Modbus RTU, isolated |
| Power supply | DC 9~24 V, about 0.2 W |
| Protection / pressure | IP68; 0~1 bar |
| Housing / size | Stainless steel 316 and POM; Φ37 mm x 261 mm |
| Cable / weight | 10 m standard, customizable; approx. 500 g |
| Installation | Side wall, top flange, flow-through, submerged, pipeline |
The same sensor body is supplied for five mounting arrangements: side wall installation on tank or channel walls, top flange mounting through a cover, flow-through installation in a bypass or sample line, submerged installation on a rail or bracket, and pipeline installation threaded into a tee. A 10 m cable is standard and can be customised, and the installation angle should stay within ±30° of vertical.
The DPH890 user manual, register map and dimension drawing are released for evaluation so an integrator can confirm wiring, protocol and mounting before the panel is designed. Contact the manufacturer's engineers to request the documentation set and a quotation.